This website is using cookies We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive all cookies on this website.

Cooling Circulators

Recirculating Chillers

Cooling devices for Dissolvers and Bead Mills

Repeatable results in laboratory, pilot plant and production

According to Winkler Model, success of a dispersion or fine grinding process is not only a function of time but more importantly a function of "Energy Density" of the system.

As the Energy density is correlated with Mechanical Power Input it is also dependent on Moment.

P = 2 π n M
 P=power [Nm/s = J/s = W] 
 π=3.141... 
 n=speed [1/s] 
 M=torque [Nm] 

The torque of a certain millbase is relevant to the viscosity which can change dramatically with the change of temperature.

Due to this, result of a trial of same formulation with same constant rotation speed in same time in wintertime with a cooling fluid of 13°C , may be completely different from the result of the summertime trial where the cooling fluid temperature is 23°C

In order get repeatable results with a constant speed operation, it is essential to work with same cooling fluid temperatures. A recirculating chiller is recommended where repeatability is important. Usage of such recirculator also prevents loss of high amount of clean water where tap water is only alternative for cooling.

Chiller
KTR
KTC
KTD
KT1
K1
K3
K4
K6
K9
Cooling Capacity
(@17°C)
480 watts
480 watts
480 watts
900 watts
1750 watts
3200 watts
4500 watts
6000 watts
9000 watts
Temperature
Fixed temperature operation @ 10°C
Fixed temperature operation @ 10°C
+4° - +35°C PID control
+4° - +35°C PID control
+4° - +35°C PID control
+4° - +35°C PID control
+4° - +35°C PID control
+4° - +35°C PID control
+4° - +35°C PID control
Temperature Stability
+/-0,2°C
+/-0,2°C
+/-0,1°C
+/-0,1°C
+/-0,1°C
+/-0,1°C
+/-0,1°C
+/-0,1°C
+/-0,1°C
Digital Temperature Display
-
Actual
Actual & Set value
Actual & Set value
Actual & Set value
Actual & Set value
Actual & Set value
Actual & Set value
Actual & Set value
System Volume
1,5 litres
1,5 litres
1,5 litres
2 litres
2 litres
4 litres
7 litres
7 litres
7 litres
Low fluid level alarm
-
Audio/ Visual
Audio/ Visual
Visual
Visual
Visual
Standard - Visual with pump interlock
Standard - Visual with pump interlock
Standard - Visual with pump interlock
Noise Level
@ 1 meters
55 dB(A)
55 dB(A)
55 dB(A)
58 dB(A)
68 dB(A)
68 dB(A)
68 dB(A)
68 dB(A)
85 dB(A)
Dimensions
E x B x Y
330 x 315 x 515 mm
330 x 315 x 515 mm
330 x 315 x 515 mm
370 x 450 x 540 mm
545 x 420 x 575 mm
540 x 550 x 713 mm
775 x 510 x 850 mm
775 x 510 x 850 mm
775 x 510 x 850 mm
Weight
30 kg
30 kg
30 kg
55 kg
66 kg
82 kg
125 kg
125 kg
125 kg

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.